ASTM D93 Pensky-Martens Intelligent Opening Flash Point Tester
1.Overview
PCK208 type intelligent opening flash point automatic tester is in
accordance with the national standard of the People's Republic of
China - Determination of flash point and ignition point of
petroleum products_ Cleveland open cup method -GB/T 3536-2008, ASTM
D92, ISO2592-2000 relevant provisions, It is designed by using
microprocessor control technology and true color touch liquid
crystal display system. This instrument is usually used for the
determination of petroleum products other than fuel oil (which is
usually determined in accordance with GB/T 261), and the flash
point of the cup is higher than 79 ° C
2.Characteristics
Chinese and English bilingual menu, true color GUI touch screen
interface design, intuitive display, simple operation (under
certain conditions can be completely familiar with the operation
process without looking at the manual), reasonable structure of the
instrument, wide test range, repeatability and authenticity. The
test process of the instrument is fully automatic. The temperature,
time and other parameters involved are displayed by the LCD. On
this LCD screen, the user will also see the working state of the
instrument and the relevant operation tips.
You only need to preset the expected flash point or initial
temperature of the substance to be tested, and the instrument can
heat up, sweep, flash point detection, detection device lifting,
air cooling, and print out the results at the specified rate,
eliminating the complex operation process that you have to do by
hand in the past. It has the function of automatic detection and
manual input of atmospheric pressure. When the automatic detection
state is valid, the manual input is invalid, and vice versa. The
instrument will automatically correct the test results according to
the atmospheric pressure changes.
The instrument has powerful data processing functions: it can
remember the results of the past 100 times of analysis, so that you
can easily query and print the historical records. It can remember
the preset parameters before shutting down each time. The
instrument comes with a hundred year calendar, which can display
the year, month, day, hour, minute and second. Optional 232
computer management interface (a set of equipment computer
software)
3.Structure of instrument
The instrument is single design, beautiful shape, compact
structure, light weight, small volume.
The control core is equipped with power supply, interface, clock,
drive, LCD display system and other functional circuits, as well as
heating electric furnace, cover opening mechanism, testing
mechanism, air source control valve and lifting mechanism and other
executive parts to complete the automatic control of the whole
inspection process.
Executive standard | GB/T3536-2008 ASTM D92 ISO2592-2000 |
Temperature measuring element | Platinum resistance (Pt100) |
Temperature range | 79ºC---400ºC |
resolution | 0.1ºC |
Repeatability of measurement | Measurement repeatability When the flash point is less than or equal to 150 ° C, it is ± 2 ° C |
heating rate | The heating rate is 14 ° C-17 ° C / min before the pre-flash point
of 56 ° C, Starting at 23 °C before the pre-flash point, it starts at 5 °C-6 °C/min. |
display usage | Display mode True color touch LCD display |
operating environment | Ambient temperature 15-35 °C Ambient humidity less than 85% |
Power requirements | 220V.ac±10% 50Hz±1Hz |
Power loss | ≤600W |
maximum size | 380*290*220mm |
quality | 15Kg |
![]() |