BGD 261 ASTM D 5895 Line Drying Time Recorder
1.Description
Drying Time
When developing a process, it is important to know the exact time
it takes for a coating to dry or cure. Coating drying time is
divided into several stages. After the coating is applied, the
first stage is when the coating levels out under gravity. Once the
coating begins to cure, a thin dry film appears on the surface.
But how do you know when the coating is completely dry?
With the BGD Drying Time Recorder, operators can easily identify each stage of
the drying process.
A ball is placed in the coating and the recorder starts to move the
ball at a predetermined speed, the mathematical formula being:
distance = speed x time. As the coating dries, the ball leaves a
trail in the coating that identifies each stage of curing.
Production Line Drying Time Recorder
The stages of film drying and curing are easy to detect, but
difficult to define chemically and physically. To evaluate them
satisfactorily, it is necessary to use instruments under controlled
conditions.
BGD offers a versatile Drying Time Recorder to help quantify the
various stages of film curing and drying, providing repeatable
results and ensuring maximum efficiency.
This reliable instrument can test the drying time or gelation
behavior of many paints and coatings by applying them via our cube
applicator (BGD 203) on a 330 mm x 24 mm glass strip.
The drying time of each stage can be easily evaluated by means of a
scale (configured according to the traverse speed). The total time
for each test can be observed on the LCD screen.
1. solvent evaporation: deep pear-shaped blots
2. sol-gel transition: continuous trace
3. surface drying: interrupted trace
4. final drying time: needle no longer penetrates film
2.Technical Parameter
| Simultaneous Testing Capacity | 6 samples (time-saving) |
| Testing Speeds | 6 hrs, 12 hrs, 24 hrs, 48 hrs (adjustable for different applications) |
| Stylus Diameter | 2mm ±0.05mm (with rounded tip; includes 6 pcs) |
| Included Weights | 6 stainless steel weights (5 grams each) for drying recording |
| Certification | Calibration certificate included |
| Compliance Standards | ASTM D 5895-03, ISO 9117-4 |